Test Kits Tk 18591
The Series 18591 Test Kits are handy instruments which simulate RTD or T/C sensors, at 12 discrete Calibration Points (CPs) for the calibration of Electronic Temperature Indicators, Recorders and Controllers. In case of T/C sensors, the Test Kit also incorporates automatic CJ Compensation.
The Test Kits are designed with a precision ladder of metal film resistors and trim pots. The ladder provides the CPs directly for RTDs. In case of T/Cs, the CPs are provided by the ladder and a CJ compensation circuit fed from a precision voltage reference powered with two pen-light cells. A battery monitoring LED indicates the sufficiency of battery voltage for maintaining precision.
OUTPUT SPECIFICATIONS
Sensor Options
RTDs & T/Cs as per the Sensor Table given below
| RTDs |
T/Cs |
| Opt.Code |
Sensor Type |
Opt Code |
Sensor Type |
Opt Code |
Sensor Type |
| P |
Pt 100 |
J |
Fe/Const |
S |
Pt/Pt-Rh(10%) |
| Q |
Pt 46 |
K |
Cr/Al |
R |
Pt/Pt-Rh(13%) |
| C |
Cu 53 |
T |
Cu/Const |
|
|
| N |
Ni 100 |
E |
Cr/Const |
|
|
Standard Options
Any of the Individual standards as below for the sensor selected above
| Option N |
ISA/ANSI |
Option J |
Japanese |
| Option B |
British |
Option D |
DIN |
| Option I |
Indian |
Option R |
Russian |
RANGE OPTIONS
The Calibration Points in °C for various Sensor Options are given in the following CP table.
| Table of Sensors & Calibration Points |
| Range Option |
A |
B |
C |
D |
E |
F |
| Sensor Codes |
P,Q,C,N |
P,J,K,T |
P,Q,J,K,E |
P,J,K |
K,S,R |
S,R,B |
| |
-50 |
-50 |
-50 |
-50 |
0 |
0 |
| |
-25 |
-25 |
0 |
0 |
50 |
200 |
| |
0 |
0 |
50 |
50 |
100 |
400 |
| |
25 |
25 |
100 |
100 |
150 |
600 |
| |
50 |
50 |
150 |
150 |
200 |
800 |
| |
75 |
100 |
200 |
200 |
300 |
1000 |
| |
100 |
150 |
250 |
300 |
400 |
1100 |
| |
125 |
200 |
300 |
400 |
500 |
1200 |
| |
150 |
250 |
350 |
500 |
600 |
1300 |
| |
175 |
300 |
400 |
600 |
800 |
1400 |
| |
200 |
350 |
500 |
700 |
1000 |
1600 |
| |
225 |
400 |
600 |
800 |
1200 |
1750 |
Accuracy (1Yr)
| RTD |
: |
±0.05% of f.s. |
| T/C |
: |
±0.05% of f.s. (At 22-32 °C Ambient ) |
| CJ Error for T/Cs |
: |
±1°C max between 5 to 55 °C |
Temp Co-eff
< 0.002% of rdg per °C
Power Supply
3V DC with 2 AA size dry cells for T/C only
Low Battery
Monitored by an LED
Enclosure Size
160(L) X 70(W) X 25(H) mm
ORDERING CODE
While ordering please specify the various options as per the following ordering code
| Model No. |
Sensor Code |
Standrds Code |
Range Code |
| 18591 |
X |
X |
X |
Example
Specify 18591KNE to order Test Kit for Cr-Al Thermocouple in ANSI/ISA standard with the following calibration points : 0, 50, 150, 200, 300, 400, 500, 600, 800, 1000, 1200 °C. |