Temperature Monitors: DigiTemp
2310N (N = 0, 1, 2, Etc.)
The series DigiTemp 2310N instruments are panel-mounted
Digital Temperature Monitors designed for RTD or T/C sensors.
They are available in Industry Standard Enclosures with cost-effective
models for alarm outputs, which can be tailored by the user,
at site, to suit his HI / LO and Fail-safe/Nonfail-safe alarm
requirements.
Input Options
As per the Sensor Table below.
| Opt. Code |
RTD or T/C Type |
Inst. Range Limits °C |
Min Span |
Max Span |
| P |
Pt 100 |
-200 ~ 800 |
100 |
1000 |
| Q |
Pt 46 |
-200 ~ 800 |
100 |
1000 |
| C |
Cu 53 |
-40 ~ 150 |
100 |
200 |
| N |
Ni 100 |
-40 ~ 180 |
100 |
200 |
| J |
Fe/Const |
-50 ~ 700 |
200 |
500 |
| K |
Cr/Al |
-50 ~ 1200 |
200 |
800 |
| T |
Cu/Const |
-100 ~ 400 |
200 |
300 |
| E |
Cr/Const |
-100 ~ 800 |
200 |
400 |
| S |
Pt/Pt-Rh(10%) |
100 ~ 1700 |
700 |
1000 |
| R |
Pt/Pt-Rh(13%) |
100 ~ 1700 |
700 |
1000 |
(Note: Spans other than above are available with reduced
accuracy)
Sensor Standard: : Calibratable
in any Standard. Factory calibrated as per DIN 43760 for
RTDs and as per ISA/ANSI for T/Cs, if standard is not specified
in the order.
Operator Interface
| No of Displays |
: |
One or Two depending on Model |
| Display Type |
: |
7 Segment 3½ dgt 12 mm or 4½ dgt 8 mm RED / ORANGE LEDs.
(20/25 mm, 50 mm LEDs and Bargraph available in Model
23104 in specific sizes) |
| Accuracy (1Yr) |
: |
| RTDs |
: |
±0.2% of rdg ±1 dgt. |
| T/Cs |
: |
±0.5% of rdg ±1 dgt ±1°C
(CJ Error ) |
|
| U/R & O/R Ind |
: |
LEDS indicate Under-range and Over-range of input parameter |
| Alarm Status |
: |
Active alarm is indicated by lighting an LED for each
alarm |
Ordering Information
Available Models
The Models and their Options are listed below.
| Model No |
I/P
Ch |
Display
Size, mm |
I/P
Opts |
Output
Options |
Power
Options |
Encl
Options |
Encl Depth |
| DigiTemp 23100 |
1 |
3½, 12 mm |
All |
I,A |
A,B,C,D |
C,D,E,F |
150 mm |
| DigiTemp 23101 |
1 |
3½, 12 mm |
All |
I,A,B |
A,B,C,D |
C,D,E,F |
200 mm |
| DigiTemp 23102 |
2 |
3½, 12 mm |
All |
I,A,B |
A,B,C,D |
D,E,F |
250 mm |
| DigiTemp 23103 |
1 |
4½, 8 mm |
All |
I,A,B |
A,B,C,D |
E |
200 mm |
| DigiTemp 23104 |
1, 2 |
Custom |
All |
I,A,B,C |
A,B,C,D |
Any |
250 mm |
Ordering Code
To Order an instrument of your choice, first identify the
Models which satisfy your alarm function requirements, then
select a specific Model on the basis of Enclosure, No of
Sensors etc. You can completely specify your choice by giving
an ordering code in the following format.
| Model No |
Input
Code |
Output
Code |
P.S.
Code |
Encl
Code |
Range Value1 |
Custom
Feature2 |
| R310N |
X |
X |
X |
X |
/NNN-MMMM/ |
G |
- Numerals in /NNN-MMMM / representing
range are a must.
- The Custom Feature is absent for standard options
in the specifications given above. For Custom requirements
the letter ‘G’ is to be appended to the Ordering Code
and deviations in any specification or option are emphasised
by ’G' against the corresponding Option Code.
|
Example 1:
Specify 23101PBAE/0-800/ to order DigiTemp 23101 for one
PT-100 RTD with a range of 0-800°C, 12mm LED Display,
two HI / LO Alarms, operating on 230V AC, 50Hz and housed
in a 72 (W) x 144 (H) x 200 (D) mm Enclosure.
Example 2:
Specify 23104PBAG/0-800/G {where G = Enclosure Cutout of
210 (W) x 115 (H), depth = 250 mm and a display of 25 mm
LEDs} to order DigiTemp 23101 for one PT-100 RTD with a
range of 0-800°C, 25 mm LED display, two HI / LO Alarms,
operating on 230V AC, 50Hz in an Encl of 210 (W) x 115
(H) x 250 (D) mm size. |
|